Customization: | Available |
---|---|
Type: | Infusion Pump |
Material: | Plastic |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specifications and functions | ||
Dimension | 146 x 193 x 95mm | |
Infusion set | Applicable to GB 8368-2005 disposals; diametersΦ3.5-4.5mm, thickness 0.8-1.2mm | |
Infusion mode | Rate mode, Time mode, Body weight mode, Sequential mode | |
Flow rate range | 0.1~1500ml/h | |
Accumulated volume | 0.1~9999 ml | |
VTBI | 0.1~999.9 ml, increment is 0.1 ml; 1000-9999ml, increment is 1ml | |
Increment | 0.1~999.9 ml/h, increment is 0.1 ml/h; 1000~1500 ml/h, increment is 1 ml/h | |
Accuracy | ≤ ±5% | |
KVO rate | 0.1-5.0 ml/h, adjustable, increment is 0.1ml/h | |
Bolus rate | 0.10-1500ml/h | |
Purge rate | 800ml/h | |
Occlusion | Low 150±125 mmHg (20kPa); Medium 525±125 mmHg (70kPa); | |
High 900±180 mmHg (120 kPa) | ||
Air bubble detection | Single bubble detection: 50, 100, 250, 500, 800l μl | |
Accumulated bubble detection: 0.1-4.0ml/h, adjustable | ||
Screen | 3.0'' LCD monochrome screen, 240 x 128 pixels | |
Titration | Yes, increment is 0.1ml/h | |
Keypad lock | Yes | |
Pause | Yes | |
Self-test system | Yes | |
Last therapy parameter memory | Yes | |
Drug library | 200 drugs | |
History record | 1500 events | |
Weight | <1.6kg (not including clamp) | |
Handle | Embedded handle | |
Mounting type | Standard: Vertical Pole | |
Alarm Volume | 1-8 levels selectable | |
Language | 8, Chinese, English, Spanish, French, Russian, Turkish, Portuguese, German | |
Alarm | ||
VTBI done, VTBI near done, KVO finish, Bottle empty, door open, IV set not | ||
installed, occlusion, air in line, drop error, system error, system abnormal, | ||
battery low, battery empty, reminder, AC power disconnection, standby end | ||
Safety Requirement | ||
AC power | 100-240V (50/60Hz, 0.4-0.14A) | |
DC power | 10-16V (2.25-1.5A) | |
Battery type | Lithium | |
Battery operation time | >4 hours@25ml/h | |
Charging time | ≤6 hours@100% | |
Classification | CF, Class I | |
IP Grade | IP 34 | |
Temperature | 5~40ºC for operating, -20~60ºC for storage | |
Humidity | 15~95% for operating, 10~95% for storage | |
Air pressure | 57~106kPa for operating, 50~106kPa for storage | |
Interface | ||
Data interface | Supported, RS232 | |
Nurse Call | Supported | |
DC input | Supported | |