Infant Ventilator PA-700B
Specifications:
1.Working Mode (Ventilation Mode): Controlled Mechanical Ventilation (CMV),
Intermittent Mandatory Ventilation (IMV),
Continuous Positive Airway Pressure (CPAP)
Manual Ventilation (Manual) ;
2.Breath rate:1 ~ 120BPM;
3.Inspiratory to Expiratory Ratio:1:1 ~ 1:3(Special order I:E range: 10:1 ~ 1:9.9);
4.Inspiratory flow:3 L/min ~ 20 L/min;
5.Tidal volume:Minimum 10 mL ~ 300 mL;
6.Minute Volume: Max. ventilation volume greater than 5 L/min;
7.Oxygen Concentration of Output Gas: 21% ~ 100%;
8.Ventilator System Compliance :not greater than 3×10-2 mL/Pa;
9.PEEP: 0 kPa ~ 2.0 kPa;
10.Inhalation Safety Valve : Safety valve shall be opened when the airway pressure is within -300 Pa ~ -800 Pa;
11.Inspiratory Pressure:1 kPa ~ 9 kPa;
12.Inspiratory Time:0.2 s ~ 3.0 s;
13.Synchronous Trigger: Pressure Trigger: -1.0 ~ 0kPa; (Special Order)
Flow Trigger: 0 ~ 10L;(Special Order)
14.Maximum Safe Pressure:not greater than 10 kPa;
15.CPAP: 0 kPa ~ 2.0 kPa;
16.Monitoring Tidal Volume:Real-time monitoring within the adjustment range of tidal volume: <100 mL:±10 mL; ≥100 mL: ±15 %;
17.Monitoring Oxygen Concentration:21% ~ 100%;
18.Power Failure Alarm : audible alarm sounds immediately in case of power outage and lasts at least for 120 s;
19.Gas Source Failure Alarm : audible and visual alarm sounds within 6 s after gas source interruption occurs and lasts at least for 120 s;
20.Airway Pressure Lower Limit Alarm Range: PEEP +0.2 kPa;
21.Airway Pressure Upper Limit Alarm Range: inspiratory pressure +0.4 kPa.