Overall Performance Equipment type Fully automatic discrete, STAT priority Analysis rate Constant speed 200T/H (without ISE) Test principle Colorimetry, turbidimetry Analysis method End-point, 2 kinetics, fixed-time, etc. support single/double wavelength, linear and non-linear calibration
Sample System Sample Tray 42 sample positions, random access of all samples Sample Volume 2-50ul, 0.1ul increment Sample probe Liquid level detection, the remaining detection, collision protection etc. Sample Containers Sample cups.
Reagent System Reagent Tray 25 reagent positions Reagent Volume R1:25-300ul, R2:10-150ul, 0.1ul increment Reagent Probe Liquid level detection, the remaining detection, collision protection etc. Reagent Refrigeration 4-15ºC, independent power supply, 24-hour non-stop refrigeration
Reaction System Reaction Tray Rotating tray, 120 cuvettes Cuvette Reusable high quality cuvette, Optical diameter 6mm Reaction Volume 180-500ul Incubation system Bi-directional liquid-air device Temperature 37ºC±0.1ºC with real-time display
Optical System Light source 12V35W halogen lamps, over 2000 hours Photoelectron road after spectrophotometry Wavelength 340,405,450,510,546,578,630,700nm Absorbance range 0-3.500Abs Resolution 0.0001Abs
Operating System PC operating system Windows XP or above Analysis control software English version graphical operating software Report printing Report formats support the user-defined mode, QC and state information etc. System connection RS-232