Baby Nurse Neonate Incubator Baby Infant Incubator

Product Details
Customization: Available
Type: Infant Incubator
Theory: Electrotherapeutics
Manufacturer/Factory, Trading Company

360° Virtual Tour

Diamond Member Since 2009

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Plant Area
101~500 square meters
Management System Certification
ISO 9001, ISO 13485
  • Baby Nurse Neonate Incubator Baby Infant Incubator
  • Baby Nurse Neonate Incubator Baby Infant Incubator
  • Baby Nurse Neonate Incubator Baby Infant Incubator
  • Baby Nurse Neonate Incubator Baby Infant Incubator
  • Baby Nurse Neonate Incubator Baby Infant Incubator
  • Baby Nurse Neonate Incubator Baby Infant Incubator
Find Similar Products

Basic Info.

Model NO.
BI-800
Group
Baby
Air Temperature
20~37
Temperature Rise Time
≤ 30min
Infant Bed Max Load
10kg
Transport Package
Wood Pakcing
Specification
ISO CE
Trademark
SEEUCO
Origin
China
Production Capacity
10000PCS/Year

Product Description

Servo Control Mode: Air temp;
Instant Parameter Displaying: Set temp, air temp, heat power rate;
Audible & Visual Alarms: Power failure, sensor failure, higher/lower temp, air temperature override, fan failure;
Power failure reset memory and self-test function;
Hood with dual slope: 4 hand-ports.
SPECIFICATION
POWER
Model B-800-1E: 110 / 120 VAC, 50Hz / 60Hz
Model B-800-2E: 220 / 240 VAC, 50Hz / 60Hz
Power Input: ≤ 400VA
Refer to IEC60601-1, this equipment belongs to ClassI, Type BF. Its operation is continuous.

ALARM
Power Alarm
Power Failure: Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle.
Fan Alarm
Fan Failure: Activates if Fan motor malfunctions.
Sensor Alarm
Air Sensor Failure: Air Sensor malfunction, refer unit to service.
Deviation Alarm
High Air Temperature: Activated if displayed Air Temperature is 1.5 0C above set point.
Low Air Temperature: Activated if displayed Air Temperature is 2.5
Below set point.
Over Temperature Alarm
Over Air Temperature: Activated when displayed Air Temperature was approach 38 in Air Mode.

System Alarm
System Failure: Activated if Temperature Control System malfunctions.
TEMPERATURE CONTROL RANGE AND CORRELATIVE SPECIFICATIONS
Air Temperature: 20~37
Temperature Rise Time: ≤ 30min
Temperature Variability: ± 0.5
Temperature Uniformity: ≤ 1
ENVIRONMENT REQUIREMENT
Environment Temperature
Operation Temperature Range: +20 ~ +30
Store Temperature Range: -40~ +55
Environment Humidity
Operation Humidity Range: 30%~75%
Store Humidity Range: 10%~100%
Atmospheric Pressure Range: 700hPa~1060hPa
OTHER SPECIFICATION
Noise Level Within Hood Environment: ≤ 55dB(Environment noise below 45dB)
Carbon Dioxide (CO2) Level within the Hood: Less than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress.

Air Velocity Over Mattress: Less than 0.35m/s
Infant Bed Max Load: 10kg
Transfusion Shelf Max Load: 2kg
FACTORY DEFAULT SETTINGS
Temperature Control Mode: Air Temperature Mode
Air Set Temperature: 32
Temperature Units:
NOTE: Open Access Panel or Access Door or the use of supplies or other equipment within the Incubator, which can alter the air flow pattern, may effect temperature uniformity, temperature variability, the correlation of the Incubator temperature reading to center mattress temperature and skin temperature.
SPECIFICATION
POWER
Model B-800-1E: 110 / 120 Vac, 50Hz / 60Hz
Model B-800-2E: 220 / 240 Vac, 50Hz / 60Hz
Power Input: ≤ 400VA
Refer to IEC60601-1, this equipment belongs to ClassI, Type BF. Its operation is continuous.
 
ALARM
Power Alarm
Power Failure: Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle.
Fan Alarm
Fan Failure: Activates if Fan motor malfunctions.
Sensor Alarm
Air Sensor Failure: Air Sensor malfunction, refer unit to service.
Deviation Alarm
High Air Temperature: Activated if displayed Air Temperature is 1.5 0C above set point.
Low Air Temperature: Activated if displayed Air Temperature is 2.5
Below set point.
Over Temperature Alarm
Over Air Temperature: Activated when displayed Air Temperature was approach 38 in Air Mode.
 
System Alarm
System Failure: Activated if Temperature Control System malfunctions.
TEMPERATURE CONTROL RANGE AND CORRELATIVE SPECIFICATIONS
Baby Nurse Neonate Incubator Baby Infant IncubatorBaby Nurse Neonate Incubator Baby Infant IncubatorBaby Nurse Neonate Incubator Baby Infant IncubatorBaby Nurse Neonate Incubator Baby Infant Incubator

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier
Product Groups
More