TECHNIQUES SPECIFICATIONS | |
Physical specifications | |
Screen: | 10.4 '' LCD Display Screen |
Suitable | Adult & Child |
Mode: | pneumatically powered and electrically controlled system |
Working Mode: | Closed; Semi-Closed; Semi-Open |
Circuit | Breathing circuit integrated standards |
Flowmeters: | 5 Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min; Air: 0.1~10L / Min |
Trolley: | Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions |
Drawer unit | TWO drawers totally extractable |
Technical Specifications: | |
Gas requirement: | Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ; NO2: 0.32 MPa to 0.6 MPa. |
safety valve | ≤12.5 kPa |
Respiratory Rate | 1~100bpm |
oxygen concentration in the mixed gas N2O/O2 |
> 21% |
Oxygen Flush: | 25~75 L/min |
modes of ventilation | A/C, IPPV, SIPPV, IMV, SIMV, PCV, VCV, PEEP, MANUAL, SIGH |
Compliance of the lung | ≤30 Ml/kPa |
PEEP: | 0 ~ 2.0 kPa |
Inspiratory trigger pressure | -1.0kPa ~ 2.0 kPa |
IMV frequency: | 1~20 times/Min |
I/E Ratio: | 8:1 ~ 1:10, Has the inverse ratio ventilation |
Tidal Volume | Adjustable range: 0 ~ 1500 ML, Display range:0~2000ML |
Inspiratory Plateau: | 0~1s |
O2 Concentration: | 21%~100% |
Waveform display: | P----T , F----T , P----V , F---V |
Sigh: | one deep breath per 70 ~ 120 controlled respirations, the inspiration time is 1.5 times of the setting point |
Maximum safety pressure: | ≤ 12.5 kPa |
Pressure limit range: | 0 ~ 6.0 kPa |
Airway pressure alarm: Audible and visual and with yellow and red color indicating |
Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa |
±0.2 kPa | |
Tidal volume alarm: Audible and visual and with yellow and red color indicating |
upper alarm: 50 to 2000ml, lower alarm: 0~1800ml |
Oxygen concentration alarm: Audible and visual and with yellow and red color indicating |
upper alarm: 21% ~100%; lower alarm:10%~80% |
Power Supply Alarm | Ac/dc power supply are after failing to send out alarm immediately Alarm Time: keep >120s |
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator. |
|
Operating conditions | |
Ambient temperature: | 10 ~ 40oC |
Relative humidity: | no higher than 80 % |
Atmospheric pressure: | 860 hPa ~ 1060 hPa |
Power requirement: | 110~220V, 50/60 Hz; |
Attention: the AC power supply used for the anesthesia machine must be well grounded. | |
Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System. |
|
Vaporizer | |
Vaporizer | 2 pcs |
Setting Range: 0.5%~5% | |
Storage | |
Ambient temperature: | -15oC ~ +50oC |
Relative humidity: | no higher than 95 % |
Atmospheric pressure: | 86 kPa ~ 106 kPa. |
It should be stored in a room without corrosive gas and well ventilated | |
Package | |
packaging box | comply with the requirement of GB/T 15464 |
Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation |
|
Moisture protection and rain protection to ensure that product is protected from natural damage. |