Detailed Photos
Features,
1, Air temp, and skin temp, servo-controlled by computer
2, Various and self-check alarms
3, Free-step mattress tilting adjustment
4, Removable humidity trough, easy to clean
5, >37ºC temperature setting
6, With the transfusion shelf and tray
7, Added with guardrail
8, RS-232 connector
9, Phototherapy unit (option)
Product Description
Function features:
1. Microprocessor based servo controlled temperature system
2. Air temperature, skin temperature, humidity, timer and heat power separately displayed by LED
3. Alarms functions: Power failure; Over temperature; Temperature deviation; Temperature sensor failure; Fan
failure;
4. Silence/reset button for alarms
5. Integral timer to record therapy time
6. Humidity is adjustable continuously
7. >37 °C temperature setting
8. A keypad lock prevent inadvertent changes the setting
9. Trough for pulling and pushing, which is convenient to add water and wash
10. Fan with low noise and long life-span
11. Temperature deviation and over temperature could be amended in front panel directly
12. A second thermal cut-off function for more safety
13. Skin temperature sensor failure protect function avoid over temperature
14. The infant bed inclination angle is adjustable
15. Six operating windows, two sides operating windows can be swiveled
16. Heater is electrically cut-out if air temperature inside incubator exceed 40 °C
17. Cabinet with three drawers, silence castors
18. Large size and double wall hood
Product Parameters
Specification,
Model parameter |
BI-5GB |
Electrical requirement |
~220V 50Hz or ~230V 60Hz |
Power consumption |
≤450VA |
Air mode temp. range |
25.0ºC~37.0ºC |
Skin mode temp. range |
25.0ºC~37.0ºC |
Air over temp. alarm |
< 38.0ºC |
Air mode high temp. range |
37.1ºC~39.0ºC |
Air high temp. range |
37.1ºC~38.0ºC |
High temp. mode alarm |
< 40.0ºC |
Air mode display temp. range |
20.0ºC~45.0ºC |
Skin mode display temp.range |
20.0ºC~45.0ºC |
Air mode alarm |
+1.5ºC or -2.5ºC |
Skin mode alarm |
±1.0ºC |
Skin sensor accuracy |
<=±0.3ºC |
Air control accuracy |
<=0.5ºC |
Mattress temp.uniformity |
<=0.8ºC |
Noise |
<=55dB(A) |