AED-7000 Medical Equipment Supply Price Of Quality Products Portable Defibrillator
Specification:
CategorySpecification
OutputBiphasic Truncated exponential
Energy Sequence150, 150, 200J
Charge Time8 sec. To 150J
10 sec. To 200J(New battery)
Analysis Time9 sec.
The maximum time from the inition of rhythm analysis to readiness for discharge with a new batteryLess than 30 seconds
The maximum time from the inition of rhythm analysis to readiness for discharge after 6 shocksLess than 35 seconds
The maximum time from initially switching power on to readiness for discharge. Less than 25 seconds
Audible Prompts20 audible prompts
Visual PromptsLED prompts
ControlsTwo buttons-On/Off, Shock
Output Energy Accuracy± 15% into any impedance from 25 to 175Ω Ohm
Output disabled when the patient impedance is outside limits20Ω To 200Ω
Battery
Part No Non-rechargeableCR123A-4x2; 12V d. C. 3.0Ah
Capacity100 discharges at 200 Joules or 120 discharges at 150 Joules
Shelf Life(25° C± 15° C)10 years(5 years storage + 5 years standby); 5years standby(after installation)
Physical
Operating Temperature0° C to 40° C
Operating HumidityRelative humidity between 30% and 95% (noncondensing)
Storage Temperature(without battery)minus 20° C to 55° C
Storge Humidity(without battery)Up to 93%(non-condensing)
Output |
Biphasic Truncated Exponential |
Energy Sequence |
150,150,200j |
Charge Time |
8 Sec.to 150j |
Analysis Time |
9 Sec |
Operating Temperature |
0°c to 40°c |