BI-200 Transport Infant Incubator
Specifications
Product Introduction
Two temperature control modes: Air mode and Baby mode controlled by micro-computer;
Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately;
Failure alarm indication;
Can be equipped with >37ºC temperature set function;
Double wall Acrylic glass hood with side door so that the bassinet can be pulled out from it;
Natural air flow humidity;
The height of whole unit can be adjusted;
Oxygen supply system;
Observation lamp.
Basic Configuration
Main body (including the hood, controller, bassinet, internal battery, and observation lamp), oxygen cylinders, oxygen supply system, skin temperature sensor, I.V. Pole, mattress, adjustable stand).
Main Parameters
AC power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 400VA
DC Power requirement: DC12V/10A or DC24V/6A
Mode of control: Air mode and baby mode controlled by micro-computer
Air temperature control range: 25ºC-37ºC(Override mode 37-38ºC)
Baby temperature control range: 34ºC-37ºC(Override mode 37-37.5ºC)
Skin temperature sensor accuracy: ±0.2ºC
The temperature difference between incubator temperature and average incubator temperature(under steady temperature condition): ≤1.0ºC
Temperature uniformity: ≤1.5ºC(mattress is on the horizontal position)
Noise inside hood: ≤52 dB(A)
Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm and so on.
Battery continuous working time: 90min(Set temperature is 36ºC-Ambient temperature is 15ºC)
Environmental temperature:
Operating range: 10 ºC~30ºC
(Set temperature should be 3ºC higher than the environment temperature)
Ambient air movement rate: >0.3 m/s,<1.0m/s
Product Packaging:
Package: each unit is packed into two cartons
Dimension (Main body): 1090mm x 655mm x 760mm
(Cabinet): 1380mm x 610mm x 400mm
Gross Weight (Main body): 80Kg
(Cabinet): 53Kg