Component |
Detail |
Power Output |
25kW |
Inverter Frequency |
40kHz |
X-ray Tube |
Rotary anode, Dual-focus, Large focus: 1.3mm*1.3mm, |
Small focus: 0.6 mm*0.6mm, Thermal capacity:900kJ |
Rotary Anode Speed |
3000rpm |
Tube Current |
200mA |
Tube Voltage |
40-125kV |
mAs |
0.4-360mAs |
Table Radiography |
Grid density: 103L/INCH;Grid ratio:8:1;Focusing distance: 120cm;Stationary type:15"×18"(import) |
Bucky Stand Radiography |
Grid density: 103L/INCH;Grid ratio:10:1;Focusing distance: 150cm;Stationary type:18"×18"(import) |
Power Supply |
380V 50Hz, Capacity: ≥30kVA |
Operation Method |
Color LCD touch screen graphical control console |
20 meters wireless remote exposure control |
Configuration Characteristics |
Table Size |
2000×760mm |
Table Height |
≤700mm |
Transverse Travel of Table |
±110mm (Electromagnetic brake ) |
Longitudinal Travel of Table |
±325mm(Electromagnetic brake ) |
Cassette Holder Grid Movement |
≥560mm( Cassette frame which can linkage with column to take face film; not linkage with the column , to take angle film.) |
X-ray Tube Rotation |
±90°(Electromagnetic brake ) |
Floor Stand Rotation |
360° ( 4×90° , Mechanical brake ) |
Axial Rotation |
0-35°( Mechanical brake ) |
SID |
420-1200mm(Electromagnetic brake ) |
Floor Stand Longitudinal Travel (along table) |
≥1350mm(Electromagnetic brake ) |
Beam Aperture Opening Duration |
Approx 30s |
Film Size |
5"×7"-14"×17"(Table radiography) |
5"×7"-17"×17"(Bucky stand radiography) |
Movement Range of Chest X-ray Frame |
560-1625mm |