Customization: | Available |
---|---|
Type: | Baby Delivery |
Function: | Diminish Inflammation |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Servo Control Mode: air temp; |
Instant Parameter Displaying: set temp, air temp, heat power rate; |
Audible & Visual Alarms: power failure, sensor failure, higher/lower temp, air temperature override, fan failure; |
Power failure reset memory and self-test function; |
Hood with dual slope: 4 hand-ports. |
Specifications of hood |
BI-3000 |
Mattress to hood |
450 mm |
Observation side of hood |
Two incline |
Operate port |
6 (2 Iris ports) |
Tubing grommets |
4 |
Front access panel |
826*310 mm |
Mattress Tray Size |
630*360 mm |
Power requirement |
≤400VA |
Temperature control mode |
Air and skin servo control |
Air Temp Control Range |
20°C 37°C |
Temperature override mode to |
20°C 39°C |
Skin Temp Control Range |
30°C 37°C |
Temperature override mode to |
30°C 38°C |
Skin Temp sensor accuracy |
≤0.3°C |
Humidity display range |
5% 99% |
Temperature fluctuation |
≤0.5°C |
Mattress temperature uniformity |
≤0.8°C |
Temperature rise time |
≤30 mim |
Noise level within hood environment |
≤55db |
Power failure alarm |
Yes |
Fan failure alarm |
Yes |
Sensor failure alarm |
Yes |
High air temperature alarm |
+1.5°C |
Low air temperature alarm |
-2.5°C |
High skin temperature alarm |
+1.0°C |
Low skin temperature alarm |
-1.0°C |
Over air temperature alarm |
≤38.0°C; ≤40.0°C |
Over skin temperature alarm |
≤38.0°C; ≤39.0°C |
System failure alarm |
Yes |
Alarm silence function |
Yes |
Environment temperature |
+20°C +30°C |
Environment relative humidity |
5% 99% |
Environment air velocity of flow |
<0.3m/s |
RS-232 Connector |
Yes |
Rail |
Yes |
Infant bed tilt adjusting |
±10° step-less |
Mattress tray Max. load |
10 kg |
Transfusion pole Max. load |
2 kg |
Cabinet Max. load |
6 kg |
Packing size (2 cartors) |
100*58*80 |
Gross weight |
120 kg |
Servo Control Mode: air temperature, skin temperature; |
Instant Parameter Displaying: set temp, air temp ,skin temp, internal humidity, heat power rate; |
Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air& skin temp override, fan failure, system failure; |
>37°C Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function; |
Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf; |
RS-232 Output Connector (Use for network system ). |