Customization: | Available |
---|---|
Classification: | Imaging Diagnostic Equipment |
Type: | X Ray Equipment |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Generator | Kvp Range | Radiography 40 ~ 150KVp |
mA Range | Radiography 25 ~ 630mA | |
Time Range | Radiography 1.0 ms ~ 5000ms | |
Lowest Current Time | 0.5 mAs | |
3-Phase Power Supply | 380VAC±38VAC | |
Power Frequency | 50Hz±1Hz / 60Hz±1Hz | |
Power Capacity | 60KVA | |
X-ray Tube | Rated Voltage | 150kV |
Rated Focal Spot Value | 0.6 | |
Anode Heat Storage Capacity | 2,30,000HU | |
Heat Storage of Tube Assembly | 1,350,000HU | |
Radiographic Table | Table Surface (L×W×H) | 2100×800×650 mm |
Table Surface Moving Range | Vertical: 900mm; Horizontal: 240mm | |
Table Surface Lock Mode | electromagnetic lock, braking force≥100N | |
Table Surface Lock Control | foot switch (permanent on status) | |
Grid Dimension | 356mm × 432mm ( 14"×17") | |
Filter Vibrating Mode | fixed | |
Dimension of The Film used in Filter | Max. 356mm × 432mm (14"×17") (17" film can only be used in table vertical direction | |
Source Assembly Stand | X-Ray Tube components Up-and-Down Motion Range | ≥1220mm |
Movement Range Between Tube Focus and Film | 1100mm | |
X-Ray Tube Assembly Vertical Motion Range | ≥2380mm | |
X-Ray Radiation Component Rotating Range Around Support Arm Axis | -120° ~ +120° | |
X-Ray Radiation Component Rotating Range Around the Vertical Axis of the Stand | 0° ~ ±90° and 90° mechanical positioning lock | |
X-Ray Tube Assembly Locking Force | Electromagnetic locking braking force ≥ 100N | |
Collimator | Mode of Adjustment | Manuel |
Radiation Field | 450×450mm [Max SID = 100cm]; 2×2mm [Min SID = 100cm] | |
Time Limit of the Light Source | ≤30s | |
Ion Chamber | Sensitivity Per Measuring Field | 60-80 pC/uGy |
Difference Of The Measuring Field | ≤5% | |
Dose Rate Range | 0.5-1000 uGy/s | |
Exposure Dose Range | 1ms-10s |