| Customization: | Available |
|---|---|
| Theory: | Electrotherapeutics |
| Control Modes: | Air Mode and Baby Mode Controlled by Micro-Comput |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| Power supply | AC220V-230V/50Hz or AC110-120V/50-60Hz |
| Max Power Output | 1000VA |
| Mode of control | Air mode and baby mode controlled by micro-computer |
| Air temperature control range | 25C-37C(override mode 37-39C) |
| Baby temperature control range | 34-37C(override mode 37-38C) |
| Accuracy of skin temperature sensor: | ±0.3C |
| Temperature variability | ≤0.5C |
| Temperature uniformity | ≤0.8C(mattress is on the horizontal position) |
| Inclination of the bassinet | ±12° |
| Noise inside hood | ≤55dB(A) (Environment noise≤45dB(A)) |
| Optional: with DC motor. Noise inside hood | ≤45dB(A) (Environment noise≤35dB(A)) |
| Failure alarm | over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, sensor box position alarm and system alarm,low water alarm and tank position alarm |
| Humidity display range | 0%RH~99%RH |
| Humidity control range | 0%RH~90%RH |
| Accuracy of humidity | ±10%RH |
| Indication range of oxygen concentration | 0%~99% |
| Accuracy of oxygen concentration | ±3% |