630mA High Frequency Radiography System
Generator:
Kvp Range:
Radiography 40 ~ 150KVp
MA Range: Radiography 25 ~ 630mA
Time Range:
Radiography :1.0 ms ~ 5000ms
Lowest Current Time: 0.5 mAs
3-Phase Power Supply: 380VAC± 38VAC
Power Frequency: 50Hz± 1Hz / 60Hz± 1Hz
Power Capacity: 60KVA
X-ray Tube:
Rated Voltage: 150kV
Rated Focal Spot Value: 0.6
Anode Heat Storage Capacity: 2, 30, 000HU
Heat Storage of Tube Assembly: 1, 350, 000HU
Radiographic Table:
Table Surface (L× W× H): 2100× 800× 650 mm
Table Surface Moving Range:
Vertical: 900mm;
Horizontal: 240mm
Table Surface Lock Mode: Electromagnetic lock, braking force≥ 100N
Table Surface Lock Control: Foot switch (permanent on status)
Grid Dimension: 356mm × 432mm ( 14"× 17")
Filter Vibrating Mode: Fixed
Dimension of The Film used in Filter:
Max. 356mm × 432mm (14"× 17") (17" film can only be used in table vertical direction
Source Assembly StandX-Ray Tube components Up-and-Down Motion Range≥ 1220mm
Movement Range Between Tube Focus and Film1100mm
X-Ray Tube Assembly Vertical Motion Range≥ 2380mm
X-Ray Radiation Component Rotating Range Around Support Arm Axis-120° ~ +120°
X-Ray Radiation Component Rotating Range Around the Vertical Axis of the Stand0° ~ ± 90° And 90° Mechanical positioning lock
X-Ray Tube Assembly Locking ForceElectromagnetic locking braking force ≥ 100N
Collimator Mode of Adjustment Manuel
Radiation Field:450× 450mm [Max SID = 100cm]; 2× 2mm [Min SID = 100cm]
Time Limit of the Light Source: ≤ 30s
Ion Chamber:Sensitivity Per Measuring Field60-80 pC/uGy
Difference Of The Measuring Field≤ 5%
Dose Rate Range: 0.5-1000 uGy/s
Exposure Dose Range: 1ms-10s
Generator |
Kvp Range |
Radiography 40 ~ 150KVp |
mA Range |
Radiography 25 ~ 630mA |
Time Range |
Radiography 1.0 ms ~ 5000ms |
Lowest Current Time |
0.5 mAs |
3-Phase Power Supply |
380VAC±38VAC |
Power Frequency |
50Hz±1Hz / 60Hz±1Hz |
Power Capacity |
60KVA |
X-ray Tube |
Rated Voltage |
150kV |
Rated Focal Spot Value |
0.6 |
Anode Heat Storage Capacity |
2,30,000HU |
Heat Storage of Tube Assembly |
1,350,000HU |
Radiographic Table |
Table Surface (L×W×H) |
2100×800×650 mm |
Table Surface Moving Range |
Vertical: 900mm; Horizontal: 240mm |
Table Surface Lock Mode |
electromagnetic lock, braking force≥100N |
Table Surface Lock Control |
foot switch (permanent on status) |
Grid Dimension |
356mm × 432mm ( 14"×17") |
Filter Vibrating Mode |
fixed |
Dimension of The Film used in Filter |
Max. 356mm × 432mm (14"×17") (17" film can only be used in table vertical direction |
Source Assembly Stand |
X-Ray Tube components Up-and-Down Motion Range |
≥1220mm |
Movement Range Between Tube Focus and Film |
1100mm |
X-Ray Tube Assembly Vertical Motion Range |
≥2380mm |
X-Ray Radiation Component Rotating Range Around Support Arm Axis |
-120° ~ +120° |
X-Ray Radiation Component Rotating Range Around the Vertical Axis of the Stand |
0° ~ ±90° and 90° mechanical positioning lock |
X-Ray Tube Assembly Locking Force |
Electromagnetic locking braking force ≥ 100N |
Collimator |
Mode of Adjustment |
Manuel |
Radiation Field |
450×450mm [Max SID = 100cm]; 2×2mm [Min SID = 100cm] |
Time Limit of the Light Source |
≤30s |
Ion Chamber |
Sensitivity Per Measuring Field |
60-80 pC/uGy |
Difference Of The Measuring Field |
≤5% |
Dose Rate Range |
0.5-1000 uGy/s |
Exposure Dose Range |
1ms-10s |