Perlong Medical Equipment Co., Ltd.
Jiangsu, China
Diamond Member Since 2009
Audited Supplier
Manufacturer/Factory, Trading Company
Sample Available
OEM/ODM Service
Factory Tour
Perlong Glycated Hemoglobin Hba1c Analyzer
Reference FOB Price: | US $2,500.00-3,000.00 / Piece |
---|---|
Min. Order: | 1 Piece |
Classification: | Biochemical Apparatus |
---|---|
Type: | Biochemistry Analyzer |
Certification: | ISO13485 |
Display: | 320X240 LCD Graphic Display, Displaying Real-Ti |
Sample Type: | Venous Blood (EDTA Anticoagulation), Peripheral |
Sample Volume: | 5UL Full Blood |
Basic Info
Model NO.
HB6601
Sampling Mode
25 Position Sample Turntable Position
Testing Mode
Batch Auto Test, Single Test for Emergency
Transport Package
Export Packing
Specification
ISO
Trademark
Perlong
Origin
Jiangsu, China
Production Capacity
100units/Week
Product Description
HB6601 glycated hemoglobin hba1c analyzer
Inside Heating
Durable Pump Tubing
Filter for avoiding blockage
Pressure Monitor
Inside Heating
Durable Pump Tubing
Filter for avoiding blockage
Pressure Monitor
Basic Parameters | |
Testing Method | Low pressure ion exchange liquid-phase chromatography |
Testing Item | Glycosylated hemoglobin A1c |
Testing Scope | 3% ~ 18% |
Testing Parameters | Precision (CV) ≤3%, Accuracy (V) ≤1.50% |
Testing Time | 4 minutes (reporting), Full-processing 6 minutes (including cleaning column and recovery time) |
Sample Type | Venous blood (EDTA anticoagulation), Peripheral finger blood |
Sample volume | 5μl full-blood for each of samples |
Functional Parameters | |
Photometer | 415nm LED Integral flow colorimeter |
Sampling mode | Full-auto 25-position sample turntable (20 sample positions, 2 quality control positions, 1 emergency position, 1 cleaning position, 1 zero position) |
Testing mode | Batch auto test, single test for emergency |
Calibration | Optional selection of 2 points or 4 points, optional selection of manual or automation Equipped with 2 groups of high and low level calibrators |
Thermostatic Control | Column and agent thermostatic control: 25 ± 0.5°C |
Reagent Matching | One piece of column supplied with consumption/300 pieces of testing samples |
Reagent Allocation | Each of standard agent matched with A, B, C, D eluent, calibrator, hemolytic agent, pump pipe |
Output Parameters | |
Display | 320x240 LCD Graphic Display, displaying real-time testing cure |
Printer | 58mm Thermal printer, printing testing cure and report |
Report Output | IFCC concentration value, NGSP area percentage, ADAG average glucose |
Data storage | 1000 test report (including testing cure) |
Communication interface | RS232 communication interface, connecting to HISILIS system |
Working Parameters | |
Power Supply | AC220V±22V 50Hz 65VA |
Overall size | 408mm×390mm×230mm |
Weight | 8Kg |
Working Environment | Temperature 12°C-30°C,Relative humidity ≤85% |