Specifications:
Flow O2: 0.05~1L/min; 1~10L/min
N2O: 0.05~1L/min; 1~10L/min
AIR: 0.2~1L/min; 1~12L/min
ORC: O2≥ 25%
O2 flush: 35~75L/min
Oxygen deficiency alarm and N2O cut-offWhen O2 ≤ 0.2Mpa, audible alarm occurs and lasts for at least 7 seconds
When O2 ≤ 0.14Mpa, N2O flow decreases correspondingly
When O2 ≤ 0.04Mpa, N2O flow is cut off
Integrated breathing circle systemAPL adjustment range: 0.5~7kPa
CO2 absorber capcity: 1.6L
VaporizerAmbient temperature: 15~35° C
Input flow: 0.25~15L/min
Output back pressure: -10kPa~20kPa
Output concentration: 0.2%~5%(± 0.2%) (Vol%)
Frequency6~40 bmp
Tidal volumeAdult: 100~1100mL Infant: 30~150mL
I: Eratio 1: 1~1: 9.9 adjustable
Flow O2: 0.05~1L/min; 1~10L/min
N2O: 0.05~1L/min; 1~10L/min
AIR: 0.2~1L/min; 1~12L/min
ORC: O2≥ 25%
O2 flush: 35~75L/min
Oxygen deficiency alarm and N2O cut-offWhen O2 ≤ 0.2Mpa, audible alarm occurs and lasts for at least 7 seconds
Model |
850 (Standard Model) |
850 (Advanced Model) |
Display |
8.4 inch TFT display |
10.4 inch TFT display |
Flowmeter Range |
O2: 0.1-10L/min |
N2O: 0.1-10L/min |
Air: 0.1-10L/min |
Ventilation Mode |
IPPV, SIPPV, VCV, IMV, SIMV, MANUAL, Standby |
IPPV, SIPPV, VCV, PCV, IMV,SIMV, MANUAL, Standby |
Back-up Power Supply |
At least 4 hours |
Tidal Volume |
Adjustable range: 10-1500ml |
Display range: 0-2000ml |
Respiratory Rate |
1-100bpm |
Inspiratory/Expiratory (I:E) Ratio |
8:1-1:10 (capable of inverse ratio ventilation) |
PEEP Range |
0-20cmH2O (electronically controlled) |
Inspiratory Pressure Trigger Range |
.-10-10cmH2O (electronically controlled) |
SIGH |
1 sigh breath in every 80-120 breaths |
Inhaled Oxygen Concentration Monitor |
21-100% |
SIMV Rate |
1-20bpm |
Inspiratory Plateau |
0-1 second |
Vaporizer Concentration Ranges |
0-5% |
Vaporizer Slots |
Double PA-I type slots |